Manufacturer’s Declaration – Electromagnetic Immunity
Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment
The Hemo-Force® is intended for use in the electromagnetic environment (for professional healthcare)
specified below. The customer or the user of the Hemo-Force® should assure that it is used in such an
environment.
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
LEVEL
(V/m)
(for
professional
healthcare)
385
380–
390
TETRA 400
Pulse
modulation b)
18 Hz
1,8 0,3 27 27
450
430–
470
GMRS 460,
FRS 460
FM c) ±5 kHz
deviation
1 kHz sine
2 0,3 28 28
710
704–
787
LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2 0,3 9 9745
780
810
800–
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
2 0,3 28 28870
930
1720
1700–
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band
1, 3,
4, 25; UMTS
Pulse
modulation b)
217 Hz
2 0,3 28 281845
1970
2450
2400–
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
2 0,3 28 28
5240
5100–
5800
WLAN
802.11
a/n
Pulse
modulation b)
217 Hz
0,2 0,3 9 95500
5785
NOTE 1: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna
and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted
by IEC 61000-4-3.