Temp. Range:
Operating: 14°F to 122°F
Storage: 4°F to 158°F
Humidity:
Operating: 70% RH at 68°F, 50% RH at 104°F
Storage: 95% RH at 95°F
Automatic Discharge: Capacitive circuits are automatically discharged when
the ‘Test’ push is released following an insulation
test.
Power Supply: Low voltage brushless a.c. generator
Fuses: 2 x 500 mA (F) 500 V H.B.C. 10 kA min 1
1
⁄4“ x
1
⁄4“
To check these fuses, short all three test leads together and set the rotary
selector switch on any Insulation test position. Press the ‘Test’ push and keep
it pressed while turning the generator handle. The needle should register
approximately 2 MΩ. A zero reading indicates that the ‘G’ fuse has ruptured.
An Infinity reading indicates that the +ve fuse has ruptured.
Fuse Replacement: Held in a screw type holder located in the base of the
instrument. Use a flat blade screwdriver to release the center part of the
holder, and remove the ruptured fuse. Replace with fuse(s) of the correct type,
size and rating.
Weight: 2.2 lb
Dimensions: 8
1
⁄4” (including generator handle) x 5” x 5”
Cleaning: Wipe disconnected instrument with a clean cloth
dampened with soapy water or Isopropyl Alcohol IPA)
*Relates to transient overvoltage likely to be found in fixed installation wiring.
ACCESSORIES
Supplied Part Number
User Guide 6172-382
Test lead set comprising:-
1 black, 1 red, 1 green test lead, with alligator clips 6220-436
Test Record Card (5 supplied) 6172-111
Optional
Carrying case 217740
Fuses (500mA) x 5 6121-289
Fused probe kit
FPK8 6111-287
Test lead set 6220-436
Test Record cards (Pack of 20) 6111-216
Publication - ‘A Stitch in Time’ AVTM21-P8B