AVTMTTR100-ENG Rev 6 Nov 2015
9
3
SPECIFICATIONS
Electrical
Type of Power
Re-chargeable NIMH battery pack, 3.6 V, 3800 mAh
Battery charge level indication on power up
Battery Life: Up to 15 hours of field operation
Charger Input Voltage
100-250 V ac, 50/60 Hz ±2Hz
Unit can be operated while charging
Charging time: 3.5 hrs (approximately)
Pollution Degree
TTR is designed for Pollution Degree II
Environmental Protection
Dust and shower protection to IP54
Output Test Voltage and Current
2 test voltages, auto selectable: 8V rms for testing distribution or power
transformers and PTs; 1.5V rms or 8V rms for
testing CTs.
Current: up to 100 mA, 4 digit resolution
Test Frequency
55Hz, internally generated providing a universal 50 / 60 Hz test set.