AVTMDELTA4000-EN Rev 6 April 2018
7
3
Specifications
Technical Specifications
Input Power:
100-240 V ± 10 %, 50/60 Hz, 16 A max
Output voltage:
0 to 12 kV, continuously adjustable
Test frequency range:
45-70 Hz (12 kV)
15-405 Hz (4 kV)
1-505 Hz (250 V)
Details relating to frequency range:
45Hz 350Hz - 12 kV – 100mA, continuous
45Hz 350Hz - 12 kV – 300mA, 4 minutes
45Hz – 400 Hz - 10 kV – 100mA, continuous
45Hz 400Hz - 10 kV – 300mA, 4 minutes
Max Current
(mA)
Frequenc
y (Hz)
Capacitanc
e (nF)
300
60
67
300
60
100
300
60
133
300
60
200
300
60
400
300
45
89
300
45
133
300
45
178
300
45
267
300
30
200
300
30
267
300
30
400
300
15
800
300
15
1600