These are loop resistances, therefore the resistance under test
must be subtracted from these figures. If the ‘Rc’ indicator is not
showing, the maximum error will not exceed 2%.
Max. Potential Spike Resistance: An additional error (typically 1%) will be introduced by a Potential
spike resistance of 75 kΩ. If the ‘
Rp’ indicator is not showing, the
maximum error will not exceed 2%.
Max. Output Voltage: 50 V
Display: 3
1
2 digit l.c.d. maximum reading 1999
Instrument Protection: IP54
Temperature Effect: <±0,1%/°C over the temperature range - 15°C to +55°C
Temperature Range: operating: -20°C to +45°C (0°C to +55°C for the DET5/4D)
storage: -40°C to +70°C (for the DET5/4D, without batteries)
Humidity: operating: 90% RH max. at 45°C
storage: 70% RH max. at 55°C
Flash Test: 2,3 kV a.c.
Voltage Withstand: In the event of a system fault the instrument will withstand 240 V
a.c. applied between any two terminals.