Megger DLRO 600 - Specifications; Measurement Ranges and Accuracy; Technical Parameters

Megger DLRO 600
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Warnings: Current flowing LED, Inadequate ‘C’
and ‘P’ contact, External Voltage,
Memory Full, Overheating.
Data: Transfer: Real time or batch download via
RS232 using Download Manager
(supplied free of charge).
Storage 300 sets of results and memo
Capacity: battery backed 10 years.
Memo field: 200 characters max.
Test Current: Range: 10 A to 600 A* unsmoothed d.c. in
1 A steps
Accuracy: ±2% ±2 A
Voltmeter input impedance: > 200 kΩ
Hum rejection: 10 V rms
Power Supply*: Full output current 208 to 265 V
50/60 Hz rms
Input Current: 25 A maximum
Reduced output current: Down to 100 V 50/60 Hz. rms
Temperature: Operation: -10
º
C to +50
º
C
Storage: -25
º
C to +65
º
C
Calibration: 20
º
C
Co-efficient: < 0.05% per
º
C
Humidity (max): 95% RH non-condensing
Altitude (max): 2000 m
Safety: IEC61010-1
EMC: IEC61326-1
Operational inaccuracies: Refer to www.megger.com
Dimensions: 410 x 250 x 250 mm
Weight: < 15 kg.
Measurement Ranges
Full Scale Resolution Volts (full scale) Max Test Current*
999.9 µΩ 0.1 µΩ 0.6 V 600 A
9.999 mΩ 1 µΩ 3 V 300 A
99.99 mΩ 10 µΩ 4 V 40 A
999.9 mΩ 100 µΩ 4 V 4 A
* Subject to supply voltage conditions (see below).
Accuracy is better than ±1% from 100µΩ to 100µΩ if the maximum
possible current is used. If lower currents are used, the following table
applies:
Current Error
2 A* 5.6% + 100 µΩ
5 A* 2.6% + 40 µΩ
10 A 1.6% + 20 µΩ
20 A 1.2% + 10 µΩ
50 A 0.8% + 4 µΩ
100 A 0.7% + 2 µΩ
200 A 0.7% + 1 µΩ
400 A 0.6% + 0.5 µΩ
600 A 0.6% + 0.3 µΩ
*It is not possible to select a current less than 10 A but since the open
circuit terminal voltage is somewhere between 2 V and 5 V (if input power
is 100 V to 250 V), actual test current can be as low as 2 A when loop
resistance is 1 Ω. If actual current is less than 2 A a ‘Current loop fail’ will
be reported.
Measurement: Modes: Manual, Auto, Continuous.
Test Time: < 8 seconds Manual or AUTO
mode, Refreshed every 2
second in CONTINUOUS mode
Display: Measurement: Large, high resolution backlit
liquid crystal display
SPECIFICATIONS