X.1 Unpacking the carton
Unpack the carton contents carefully. There are important documents that you should read and keep for future
reference.
Please complete the pre-paid warranty card and return it to Megger Instruments Limited as soon as possible to
help us reduce any delays in supporting you should the need assistance.
X.2 Safety Warnings
The following safety warnings and precautions must be read and understood before the instrument is
used. They must be observed during use.
! For safety, only connect the PAT to a supply that is properly earthed. If in doubt, the supply should be
checked by a qualified electrician.
! Do not use the instrument if there are any signs of damage.
! All test leads, probes and clips must be in good order, clean and with no broken or cracked insulation.
! Probes and clips should be held behind the finger guard.
! Test leads not used during a measurement should be disconnected from the appliance tester.
! For dual voltage testers, both sockets can be live simultaneously.
! Only connect one asset to the PAT during testing.
! Tests should be carried out in the order recommended below. An appliance that fails a test should be
repaired before further testing is carried out.
Recommended sequence:
1. Earth bond/ continuity of the protective earth conductor (Class I devices)
2. Insulation test (or earth leakage)
In addition further tests can be performed, but ONLY AFTER 1 & 2 above
3. Operation test
4. Leakage test
! Only perform an operational test after the earth bond and insulation tests have been completed, as this
test operates at mains voltage.
! During testing, ensure no hazard will exist as a result of normal running or under fault conditions.
! During testing the unit under test (asset) should not be touched, other than using the appropriate
accessories, as faulty appliances can present a shock hazard.
! Do not touch the exposed parts of test leads during tests as hazardous voltages may be present due to