4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
79
Config.Monitoring.Reagent.ReagCap 1...1000...9999, OFF
Config.Monitoring.Reagent.RCounter 0...9999
Config.Monitoring.Reagent.ClearCount $G
Config.Monitoring.Reagent.Drift 0...99, OFF
Monitoring of reagent live time.
.Determ: Number of determinations.
.DCounter: Counter of determinations already carried out.
.MaxTime: Maximum live time of reagent in days.
.TCounter: Time already elapsed since last reagent change.
.ReagCap: Reagent capacity in mg water.
.RCounter: Counter of reagent capacity.
.ClearCount: Clears all above counters.
.Drift: Stable drift in ug/min.
4.2.2.32. Config.Monitoring.Change $G, $S
Config.Monitoring.Change.Status auto, man., OFF
Config.Monitoring.Change.WaitTime 0...999 999
Config.Monitoring.Change.AspVol 0...100...9999
Config.Monitoring.Change.SolventVol 0...100...9999
Config.Monitoring.Change.Rinse 0...9999
Config.Monitoring.Change.NoRinse 1...9
Changing of reagent. With a connected Dosino, the reagent is changed with
&Config.Monitoring.Change $G. The nod &Config.Monitoring.change.Status
has to be OFF.
Parameters for automatic reagent change:
.WaitTime: Waiting time in s after switching off the stirrer.
.AspVol: Volume in ml of used reagent to be aspirated.
.SolventVol: Volume in ml of new reagent to be added.
.Rinse: Volume in ml of rinsing reagent.
.NoRinse: Number of rinsing cycles.
4.2.2.33. Config.Monitoring.Validation.Status ON, OFF
Config.Monitoring.Validation.Interval 1...365...9999
Config.Monitoring.Validation.Counter 0...9999
Config.Monitoring.Validation.ClearCount $G
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
4.2.2.34. Config.Monitoring.Service.Status ON, OFF
Config.Monitoring.Service.Date XXXX-XX-XX
Monitoring of service interval.
4.2.2.35. Config.Monitoring.DiagRep ON, OFF
Printing of system test report after each switching on of the Coulometer.
4.2.2.36. Config.PeriphUnit.CharSet1 Epson, Seiko, Citizen, HP, IBM