Questions and Answers

  • A
    Angelica SanchezAug 2, 2025
    Why is the drift too high in my Metrohm 756 KF?
    • A
      April MonroeAug 2, 2025
      High drift in Metrohm Measuring Instruments can be caused by moisture in the titration vessel. To resolve this, shake the titration vessel. Other causes include exhausted or contaminated reagent (exchange the reagent), exhausted molecular sieve (check the molecular sieve), a pierced septum (check the septum), faulty seals (check the seals), rough ground joint sleeves (check the ground joint sleeves), a polluted or moist generator electrode diaphragm (check the generator electrode diaphragm), or a sample matrix that consumes iodine (change the reagent more often). Also, ensure the seals are screwed tight and allow the instrument to run overnight.
  • K
    Karen HollandAug 6, 2025
    What to do if the result is too high on Metrohm 756 KF?
    • M
      Melissa Strong DVMAug 6, 2025
      If you are getting results that are too high with your Metrohm Measuring Instruments, it could be due to a titration vessel that hasn't been properly conditioned. Try shaking the vessel and wait until the drift has stabilized. Another reason may be that your sample contains substances which can be oxidized. You can also set the generator current to 400 mA and set a higher stop drift. The drift correction might be too small, especially with unstable drift or manual drift correction.
  • G
    Gina OwensAug 10, 2025
    How to fix unstable drift in Metrohm 756 KF Measuring Instruments?
    • M
      Michael MedinaAug 10, 2025
      Unstable drift in Metrohm Measuring Instruments can be addressed by ensuring efficient mixing without air bubbles by stirring and resetting the control parameters to standard values.
  • M
    Michelle DelacruzAug 15, 2025
    How to shorten titration times on Metrohm Measuring Instruments?
    • J
      James MccoyAug 15, 2025
      If titration times are too long with your Metrohm Measuring Instruments, wait until the drift during conditioning becomes stable. Also, set a higher stop drift, set the control range smaller and the max.rate higher.
  • T
    tleeAug 19, 2025
    What causes low results in Metrohm 756 KF?
    • T
      Todd WilsonAug 19, 2025
      Low results in Metrohm Measuring Instruments may occur if the drift correction is too large, the stop drift is too high, or the min.rate is too low. Another cause may be that the sample releases iodine.
  • J
    James ThomasAug 23, 2025
    Why are the results widely scattered on Metrohm 756 KF Measuring Instruments?
    • A
      Amy Price MDAug 23, 2025
      Widely scattered results with Metrohm Measuring Instruments can stem from an inhomogeneous sample or poor reproducibility of sample addition. Unstable drift may also contribute to this issue.
  • F
    fcoxAug 27, 2025
    How to correct oven parameters in Metrohm 756 KF Measuring Instruments Coulometer report?
    • A
      Alexandra YoungAug 27, 2025
      To correct wrong oven parameters in the Coulometer report of your Metrohm Measuring Instruments, switch off the report output at the oven.

Summary

Introduction to the 756;831 KF Coulometer

Parts and Controls Overview

Overview of the instrument's physical components and their functions.

The Wet Chemistry Workplace

Principle of Coulometric KF Determinations

Explains the fundamental theory behind coulometric Karl Fischer titrations.

Titration Vessel Setup Guide

Step-by-step guide for assembling the titration vessel and its components.

Optimal Working Conditions

Details optimal settings and environmental factors for accurate determinations.

Manual Operation Procedures

Understanding the Keypad

Describes the layout and function of each key on the instrument's keypad.

Configuration Settings

Explains how to configure instrument settings for various operational modes.

Parameter Adjustment

Details how to adjust titration and control parameters for optimal performance.

Data Output Options

Covers options for printing and exporting determination results and other data.

RS232 Interface Operation

General Rules for RS232 Communication

Outlines the fundamental principles and rules for controlling the instrument via RS232.

Remote Control Command Reference

Lists and describes commands for remote operation of the instrument.

RS232 Interface Technical Properties

Details the technical specifications and handshake protocols of the RS232 interface.

Troubleshooting and Error Resolution

General Troubleshooting Guide

Provides a guide to identifying and resolving common operational problems and issues.

Error and Special Message Explanations

Lists and explains specific error codes and messages displayed by the instrument.

Measuring Input Testing Procedure

Describes a procedure to test the instrument's measuring input and electrode connectivity.

Instrument Preparations and Setup

Coulometer Initial Setup

Covers the initial setup and connection of the Coulometer and its accessories.

Connecting the Coulometer to Dosino

Guide for connecting the Dosino for automatic reagent exchange.

Connecting the 768 KF Oven

Instructions for connecting the 768 KF Oven for automated sample processing.

Connecting to a PC

Details on connecting the instrument to a PC for data management.

Connecting a Remote Box

Information on connecting a remote box for input devices like barcode readers or keyboards.

Appendix and Documentation

Technical Specifications

Provides detailed technical data and specifications for the instrument.

Remote Socket Pin Assignment

Diagram and explanation of the pin configuration for the remote socket.

Coulometer Validation and GLP Mode

Describes procedures for validating the instrument's performance according to GLP standards.

User Methods and Customization

Information on creating, modifying, and using custom operational methods.

Warranty and Certificates

Information regarding product warranty, certificates of conformity, and declarations.

Scope of Delivery and Ordering

Lists the included accessories and ordering information for the instrument.

Metrohm 756 KF Specifications

General IconGeneral
BrandMetrohm
Model756 KF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals