4.2 Remote control commands
756/831 KF Coulometer, Instructions of Use
95
4.2.2.98. Setup.Initialise $G
Setup.Initialise.Select ActMeth, Config, Silo,
Assembly, Setup, All
Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Config: All values under &Config.
Silo: The silo memory is deleted. Same function as delete entire silo.
Assembly: All values under &Assembly.
Setup: All values under &Setup.
All: Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
4.2.2.99. Setup.RamInit $G
Initializes instrument, see page 110. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted.
Command only possible in the inactive state of the Coulometer.
4.2.2.100. Setup.InstrNo $G
Setup.InstrNo.Value serial number, 8 ASCII characters
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
4.2.2.101. Diagnose.Report $G
Output of the report containing the adjustment parameters. The Coulometer
has to be in its inactive basic state.
4.2.2.102. Diagnose.Simulation.Keycode 0...29
0
1
2
3
4
28
21
26
8
9
10
11
12
5
13
29
16
17
18
19
20
25
27
24
15 6 2223
Entering a keycode is like pressing the corre-
sponding key. The keys have the following key-
codes: