7.4 User methods
756/831 KF Coulometer, Instructions for Use
138
7.4.2 Working with the 774 Oven Sample Processor
When working with the 774 Oven Sample Processor an extraction period is required in
order to prevent the titration being switched off prematurely.
The RS interface of the Oven Sample Processor is connected to COM1 of the Coulometer
(cable 6.2125.110). If you do not make this connection then the parameter
oven under
preselections must be set to off (no oven data in the Coulometer report).
Determination method, parameters:
'pa
756 KF Coulometer 5.756.0010
date 1997-11-19 time 17:56 0
KFC-B 774-Det
parameters
>control parameters
EP at U 50 mV
dynamics 70 mV
max.rate max. ug/min
min.rate 15 ug/min
stop crit: rel.drift
rel.drift 5 ug/min
>titration parameters
pause 0 s
extr.time 180 s
start drift 10 ug/min
I(pol): 10 uA
electrode test: ON
temperature 25.0 °C
time interval 2 s
max.titr.time OFF s
>statistics
status: OFF
>preselections
drift corr: auto
req.ident: OFF
req.smpl size: OFF
smpl unit: g
limit smpl size: OFF
text id1 id1 or C21
text id2 id2 or C22
text id3 id3 or C23
cell: no diaph.
generator I: 400 mA
oven: COM1
activate pulse: OFF
------------