7.4 User methods
756/831 KF Coulometer, Instructions of Use
137
Blank value method, parameters:
'pa
756 KF Coulometer 5.756.0010
date 1997-11-19 time 17:51 0
BLANK Oven-Blk
parameters
>control parameters
EP at U 50 mV
dynamics 70 mV
max.rate max. ug/min
min.rate 15 ug/min
stop crit: rel.drift
rel.drift 5 ug/min
>titration parameters
pause 0 s
extr.time 300 s
start drift 20 ug/min
I(pol): 10 uA
electrode test: ON
temperature 25.0 °C
time interval 2 s
max.titr.time OFF s
>statistics
status: ON
mean n= 3
res.tab: original
>preselections
drift corr: auto
req.ident: OFF
req.smpl size: OFF
smpl unit: g
limit smpl size: OFF
text id1 id1 or C21
text id2 id2 or C22
text id3 id3 or C23
cell: no diaph.
generator I: 400 mA
oven: COM1
activate pulse: OFF
------------