7.2 Pin assignment of the "Remote" socket
756/831 KF Coulometer, Instructions of Use
129
7.2 Pin assignment of the "Remote" socket
external Function
Inputs
pin 21 (Input 0)
pin 9 (Input 1)
pin 22 (Input 2)
pin 10 (Input 3)
pin 23 (Input 4)
pin 11 (Input 5)
pin 24 (Input 6)
pin 12 (Input 7)
Start
Stop
Enter
Outputs
pin 5 (Output 0)
pin 18 (Output 1)
pin 4 (Output 2)
Ready
inactive
Conditioning ok,
active if Cond.ok
Titration,
active during titration
Functions see page 130.
t
p
> 100 ms
t
p
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
+5 V
Not used
Not used