3.15 Storing determination results and silo calculations
756/831 KF Coulometer, Instructions for Use
50
3.15 Storing determination results and silo calcula-
tions
3.15.1 Storing determination results
If the sample-specific data of the silo memory should be kept after the determination and
supplemented by results, the following entry is necessary:
In the method under <DEF>:
Assignment of the determination results to C24 and/or C25:
Assignment of determination results
The determination results are assigned in key <DEF>.
The display texts of the Coulometer are shown to the
left. The values are the default values.
>silo calculations
C24=
C25=
Silo calculations
Assignment to C24 (RSX, H2O, CXX)
Calculated results (RSX), endpoint (H2O) or variables
CXX can be stored as C24.
Same procedure for C25.
Important:
Ensure that there is still sufficient space for storing the results C24 and C25. (In the report
<PRINT><USER METH><ENTER> the number of free bytes is shown.) Result name,
value and unit are stored. The memory requirements can be estimated as follows:
Result with text (8 characters) and unit (5 characters): 32 bytes
def
>formula
>silo calculations
>common variables
>report
>mean
DEF
(
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 756 KF and is the answer not in the manual?

Metrohm 756 KF Specifications

General IconGeneral
BrandMetrohm
Model756 KF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals