3.9 Common variables
756/831 KF Coulometer, Instructions of Use
39
3.9 Common variables
Common variables are used for:
Determination of a blank value with method 1. Using this blank value in various other
methods. Mode BLANK creates the common variable C39 (default setting).
Determination of a result with method 1. Reconciliation of this result in various other
methods.
You may view the values of the common variables with <CONFIG>.
With <DEF>, results can be allocated as common
variables. The entries are specific to the method and
are stored in the method memory.
Common variables (in expert mode only):
Assigns values as common variables.
The display texts of the Coulometer are shown to the
left. The values are the default values.
>common variables
C30=
C31
C39=
Allocation for common variables
Common variable C30...C39 (RSX, H2O, CXX, MNX)
Results (RSX), endpoint (H2O), variables (CXX), and
means (MNX) can be assigned.
The values of the common variables remain in force for
all methods until they are overwritten or deleted. They
can be viewed and entered manually under the
<CONFIG> key.
Delete allocation with <CLEAR>+<ENTER>.
def
>formula
>silo calculations
>common variables
>report
>mean
DEF
(
2