4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
96
4.2.2.103. Diagnose.ScreenDump $G
The content of the 756 Screen will be dumped to the COM which is given for
manual reports (key <CONFIG>, >peripheral units). A screen dump onto the
internal printer is not possible.
4.2.2.104. Diagnose.IntPrinter.HeatTime 1...4.0...10
Diagnose.IntPrinter.MotorSpeed 2...3.0...9
Settings for the internal printer.
.HeatTime: Heating time for the dots in ms. Input in steps of 0.5 ms. Longer
heating times give darker printouts.
.MotorSpeed in ms per step (6 steps = 1 dot). Small numbers give high print-
ing speed.
If you wish to speed up the internal printer, set low heating times as a first
measure, then low motor speed.