5.1 Troubleshooting
756/831 KF Coulometer, Instructions for Use
104
Problem Possible causes and remedies
Result too low Drift correction too large, i.e. the drift was too high at the
start or unstable drift.
Stop drift too high.
Min.rate too low
Sample releases iodine.
Results widely scattered Inhomogeneous sample? Poor reproducibility of sample
addition?
Drift unstable.
Titration times too long Wait until drift during conditioning becomes stable.
Amount of water too large, see sample size guidelines on
page 10.
Set stop drift higher.
Set control range smaller, set max.rate higher.