4.18 Initial start-up
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
930 Compact IC Flex ChS/PP/Deg (2.930.1360)
4
Deaerating the high-pressure pump
Use the syringe to aspirate eluent until there are no more air
bubbles in the eluent aspiration tubing.
5
Adjusting the contact pressure of the peristaltic pump
Adjust the contact pressure of the peristaltic pump (see "Setting
the contact pressure correctly", page 26).
6
Rinsing the instrument without columns
Rinse the instrument (without columns) with eluent for 10
minutes.