Safety
4
Electrical safety
z
Before initial use of the device, assess the electromagnetic environment in
which the device will be used.
z
Do not use the device in close proximity to the source of strong
electromagnetic fields, such as unshielded sources of radiated emissions.
Radiated signals can reduce the accuracy of the results.
z
Ensure that the device is properly grounded, and the grounding resistance is
less than 4 Ω. Failure to do so might result in inaccurate experiment results
or even electric shock.
z
Do not remove the device cover and expose the inner components outside.
Otherwise, it might cause electric shock.
z
Only the power cord of the manufacturer can be used.
z
Prepare the laboratory and power supply according to
Getting started on
Page 27.
For more details, refer to the relevant site preparation guide.
z
Ensure that the input voltage meets the device requirement.
Mechanical safety
To avoid the device falling off and personal injury, place the device on a
level surface that meets the load-bearing requirements and ensure that the
device cannot be easily moved.
Components safety
Only the software that has been provided by the manufacturer can be
installed and used on the computer. Unknown software might interfere
with normal device functions, or even cause data loss. If you need to install
antivirus software, contact the technical support in advance.
z
Do not uninstall the control software by yourself. If any problem occurs
during software operation, contact the technical support.