Getting started
34
Washing cartridge and washing flow cell
Washing cartridges are delivered with the device for wash
procedures. Used sequencing flow cells function as washing flow
cells.
For how to prepare the cartridge, contact the technical support. And
for how to perform a wash manually, refer to
Performing a wash on
Page 48
.
Extra washing reagents are not provided by the manufacturer, but you can
purchase them as needed from the authorized sales representatives.
Preparing the pure water container
1. Ensure that the pure water contained is sufficient.
Insufficient pure water will result in sequencing failure.
For details about pure water consumption, refer to the relevant
reagent kit user manual. For details about maintaining the pure
water container, refer to
Maintaining the pure water container on
Page 49.
2. Connect the pure water container to the device.
1) Place the fixing plate on the lid and align the holes. Insert
the pure water tube into the pure water container through
the aligned holes until the tube reaches the bottom of the
container.
Ensure that the tube goes through the handle, as shown in the
figure below.
Figure 14 Inserting the pure water tube