other, a fourth measurement is performed before the result will be
displayed. In rare cases the blood pressure is such unstable that even
after four measurements the data vary too much. In this case, «ERR 6» is
shown and no result can be given. If one of the measurements causes
an error message,it will be repeated for one more time. In case that any
additional error occurs, the measurement will be discontinued and error
code displayed.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure
measurement for any reason (e.g. the patient feels
unwell), the O/I button can be pressed at any
time. The device then immediately lowers the
cuff-pressure automatically and enter into the
sleep mode.
5.7. Memory – storage and recall of the
measurements
The blood-pressure monitor automatically stores
the last 30 measurement values for each user. By
pressing the MEMORY button, average of existing
memory readings displayed first with an «A» and
at the lower left. Press MEMORY button again, the
last measurement (MR 30) as well as the future
last 29 measurements (MR29, MR28,…, MR 1) can
be displayed one after the other. Data obtained in
Average Mode are indicated by a symbol.
(MR1: Values of the measurement) (MR2: Values of the last measurement
before MR1)
16
MR