617
MEASUREMENT MODE
:MEAS:TEST?
Measurement test query.
Parameters:
None.
Response:
0 AC measurement type.
1 Rdc measurement type.
:MEAS:TRIGger
Trigger a measurement using the current settings.
Parameters:
None.
Response:
For AC measurements the response will be the first and second measurements separated by
a comma.
Example: 68.860E-9 , 13.0E+6
For Rdc measurements the response will be a single measurement result.
Example: 6.2295E+3
:MEAS:FREQuency <real>
Set frequency of AC measurement.
Parameters:
The required frequency in Hertz. The
unit suffix ‘Hz’ is optional.
Example: MEAS:FREQ 1k
MEAS:FREQ 1000 Hz
MEAS:FREQ 1E3
are all equivalent commands and set
the test frequency to 1kHz.
Response:
None.
:MEAS:FREQuency?
Returns the current AC test frequency.
Parameters:
None.
Response:
Returns the current test frequency in
engineering format.
Example: +.10000000E+04
for a test frequency of 1kHz.