627
MULTI-STEP MODE
:MULTI:FREQuency <real>
Set frequency for currently selected step.
Parameters:
The required frequency in Hertz. The
unit suffix ‘Hz’ is optional.
Example: MULTI:FREQ 1k
MULTI:FREQ 1000 Hz
MULTI:FREQ 1E3
are all equivalent commands and set
the test frequency to 1kHz for currently
selected step.
Response:
None.
:MULTI:FREQuency?
Frequency query of currently selected step.
Parameters:
None.
Response:
Returns the current test frequency in
engineering format.
Example: +.10000000E+04
for a test frequency of 1kHz.
:MULTI:LEVel <real>
Set drive level for currently selected step.
Parameters:
The required level in Volt. The unit
suffix ‘V’ is optional.
Example: MULTI:LEV 1.2V
MULTI:LEV 200m
will select drive levels of 1.2V and
200mV respectively.
Response:
None.
:MULTI:LEVel?
Drive level query of currently selected step.
Parameters:
None.
Response:
Returns the current test level in
engineering format.
Example: +.20000000E-01
for a test level of 20mV.