EasyManua.ls Logo

Microtest 6377 - Page 66

Default Icon
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6–28
MULTI-STEP MODE
:MULTI:SPEED <disc>
Select the required measurement speed for
currently selected step.
Parameters:
MAX Maximum speed.
FAST Fast speed.
MED Medium speed.
SLOW Slow speed.
Example: :MULTI:SPEED SLOW
will select slow speed for currently
selected step.
Response:
None.
:MULTI:SPEED?
Returns the test speed of currently selected
step.
Parameters:
None.
Response:
Returns the test speed as an integer
according to the table:
0 Maximum
1 Fast
2 Medium
3 Slow
Example: 1
indicates that Fast measurements are
selected.
:MULTI:HIgh-LIMit <real>
Set the higher test limit of the currently
selected step.
Parameters:
The required higher limit.
example: :MULTI:HI-LIM
10.0
will set a high limit to 10.
Response:
None.
:MULTI:HIgh-LIMit?
Returns the high limit value of the
currently selected step.
Parameters:
None.
Response:
The high limit value in engineering
format.
Example: +.50000000E+01
indicates a high limit to5.