5
Specifications - M9
PHYSICAL DIMENSIONS
Overall Length............................................................. 19" (483 mm)
Overall Width .............................................................. 15" (381 mm)
Overall Height ............................................................. 16" (406 mm)
Counter Area................................................................ 16 1/2" x 14"
(419 mm x 356 mm)
Chamber ........................................................................ 9" Dia x 15"
(229 mm Dia x 381 mm)
Trays
(Large) (2) ..................................................................... 12" x 8" x 1"
(305 mm x 203 mm x 25 mm)
(Small) (2) ............................................................... 12" x 5 1/2" x 1"
(305 mm x 140 mm x 25 mm)
WEIGHT
Empty.......................................................................... 73 Lb (33 Kg)
With Shipping Carton .................................................. 81 Lb (37 Kg)
WATER RESERVOIR CAPACITY APPROX. ....................... 0.88 GAL.
(3.31 Liters to Full Mark)
ELECTRICAL REQUIREMENTS
220-240 VAC, 50-60 Hz Model .............................................. 10 AMP Circuit, Single Phase
A separate (dedicated) circuit is recommended for this sterilizer. The sterilizer should not be
connected into an electrical circuit with other appliances or equipment unless the circuit is
rated for the additional load.
MAXIMUM POWER CONSUMPTION
220-240 VAC, 50-60 Hz Model .................................................. 1500 Watts, 6.3 A @ 240 V
Chamber Pressure At 270°F (132°C).............................................................31 PSI (215 kPa)
Safety Valve Setting .......................................................................................35 PSI (241 kPa)