EasyManua.ls Logo

Miele G 7782 - Indications for Use; High-Level Disinfection

Miele G 7782
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
High-level disinfection
The G 7782 Washer/Disinfector is suit
-
able for automatic treatment of medical
instruments and accessories. It cleans
and, when using the disinfection cycles
simultaneously, thermally disinfects at
200°F (93°C) instruments and accesso
-
ries.
Areas of effectiveness are physical re
-
moval or thermal inactivation of vegeta
-
tive bacteria including mycobacteria,
fungi, fungal spores, and viruses.
These categories include the Mycobac
-
terium tuberculosis and the
Enterococcus faecium.
IMPORTANT: Only the cycles labeled
‘DISINFECTION’ disinfect! Only use
these cycles when handling contami-
nated items. All other cycles are de-
signed to clean items without disinfect-
ing them!
The G 7782 Washer/Disinfector
achieves high-level disinfection as
defined by the Spaulding categoriza
-
tion.
IMPORTANT:
Guideline
for Handwashing and Hospital Environ
-
mental Control, 1985,
Instruments Suitable for Processing
in the Washer/disinfector
All instruments, accessories, and other
items to be cleaned and disinfected in
the G 7782 Washer/Disinfector must
have the following properties:
^ Heat resistance to temperatures of at
least 200°F (93°C)
^ Corrosion resistance in the presence
of heat, alkalinity and water.
Certain items and materials are not
suitable for processing in the
Washer/disinfector.
Do not process:
^
carbon steel devices as they may
corrode;
^
flexible endoscopes, light cables, fi
-
ber optics, rotating devices, motors
or other electrical equipment unless
the processing is approved by the
device manufacurer.