ETHOS EZ User manual MA073-006 Terminal chapter Page - 48 -
OK
If the program runs to the end (no manual stop or abort), then at
the end of the last heating step the
Ventilation Stop
icon will show
and at the end of the ventilation time the OK icon will be displayed.
Touching on
ok
the
Start
icon will be shown again.
Methods
Parameter Here are set the program parameters before
program start. These parameters are valid for the
entire process.
Program Here the microwave program can be edited. Each
microwave program can be made of max. 50
steps.
Info Here it is possible to enter additional information
about rack and reactor vessels.
Name Name of the opened method.
Limitations for User: User cannot create or delete methods. User – when entering the
applications – will get to a list from which he can select an existing method
.
Open new Methods
NEW METHOD
This opens a new empty method. The currently opened
method will be deleted from the main memory. Always
save before opening a new method
.