ETHOS EZ User manual MA073-006 Terminal chapter Page - 50 -
Parameters
Screenshot of Parameter Setting
Parameter settings are valid for the complete duration of the microwave program.
Visualization of parameters menu can change depending on the settings of the unit and
components/options that are installed.
Stirrer
STIRRER
This is to set the speed of the stirrer (optional).
The content of a single vessel or of several vessels contained in
the pressure reactor can be simultaneously stirred with magnetic
bars. Agitation is continuously adjustable from the control
terminal.
To operate the stirrer, always push the Test button first, and
then slowly adjust the strength with the ruler.
QP limit
QP
The QP is an optional sensor that allows to detect the vapours
concentration inside the microwave cavity. The sensitivity of
the QP sensor is set by scrolling the ruler; The lower is the set
value, the higher is the sensitivity of the sensor, and a low
vapours concentration is detected. The QP sensor is activated
checking the QP-Check” icon. If the QP-Check is activated
and the vapors concentration exceeds the preset limit value the
software cuts the microwave irradiation but the program
continues to run. As soon as the vapours concentration returns