ETHOS EZ User manual MA073-006 Terminal chapter Page - 52 -
Rotor control
ROTOR CONTROL
Rotor control must be always “ON” and it is dedicated for
infrared sensor
Door locking
DOOR LOCKING
The door is closed if the unit is switched off or until the
temperature is not below the set limit. During the cooling
step a 3°C hysteresis delay is automatically applied to the
set limit. When the door is locked, an intermittent key-icon
appear on screen.
Continuous power
CONTINUOUS POWEROTOR
If the icon is activated the system will deliver continuous power;
if it is disable the unit will work in pulsed power
Program
Program
In the
Program
section the heating program is generated or edited.
Each program can include up to 50 Steps.
The Program Steps consist of settings for Time, Power, Temperature (T1 and T2) and
Pressure.