ETHOS EZ User manual MA073-006 Terminal chapter Page - 77 -
- The temperature probe is not working; check if it reads the correct temperature or it is out
of scale.
- Pressure leakage from the reference vessel; check the complete reference vessel assembly,
see the Tips and Techniques and rotor user manuals.
- Wrong microwave program; increase the time or power, or reduce the temperature of the
microwave program.
List of Questions and (Error) Messages
The following pages describe most of the prompts/questions and error messages, ordered as follows:
Login
Nr. Text Description/Remarks
46 You entered an invalid password.
Please enter the correct password
or cancel your login.
The entered password is not valid for the selected
user.
Enter the correct password or cancel your login.
47 You have no permissions for this device.
Please contract your administrator.
The unit is in Login-Mode “with Flash Card”. The
signature of the Flash Card belongs to a user who has
no access authorization to this unit.
53 The entered username does already exist.
Usernames cannot be entered twice.
Please enter another name.
The names of the users must be different. When
entering or modifying an user, never use a name
already entered.
Sensors, Control and Twist
Nr. Text Description/Remarks
2 In your system are no sensor inputs
Configured.
Please contact your local service partner.
No sensors are configured in your system. Sensor
configuration can only be carried out by Service (see
Chapter Sensor data).
5 Your system contains a sensor for
T1 measurement. Therefore the
Curve for T1 was checked.
This curve is not sufficiently defined.
Do you really want to start the application?
The curve of a sensor is not defined or partly not
defined. This means that during run the value
measured by the sensor is not checked. This can
cause uncontrolled power emission (see also Chapter
for parameters description )
11 The temperature sensor check detected an
Error.
The value of T1 changed from 25.0°C to
100.0°C
between on second and has exceeded the limit
of 50.0°C.
The Sensor-Check detected an error in temperature
measurement (see Chapter Sensor check)
This message depends on the actual and theoretical
values and can therefore vary correspondingly.
Check the temperature sensor and positioning of the
sensor (Infrared-Sensor).
3 Your have not entered a value for the element
"Control for".
Please select on of the available sensor inputs.
You have defined no Parameter for the power control.
You have to set these parameters in the Method
Program Register. (see corresponding chapter)
24 Twist function will be disabled.
Temperature and pressure sensors
can be damaged. Are you sure to deactivate the
Security question about deactivation of the Twist-
Function (see also corresponding chapter).