ETHOS EZ User manual MA073-006 Terminal chapter Page - 78 -
Twist function?
67 Twist error
A twist error occurred.
Check the twist mode of your system.
The Twist-Logics does not give any signal of changing
direction. Check the Twist-Function (see also
corresponding chapter)
Files and Data collection
Nr. Text Description/Remarks
29 You have modified the current data collection.
Do you really want to ignore these modifications
and load the new data collection ?
You have not yet saved the current method. By loadin
g
a new method the current settings of the microwave
program will are overwritten.
22 You have modified the current data collection.
If you leave this page all data will be lost.
Do you really want to ignore these modifications
and leave this pages ?
You have not saved the current method. When you
leave the page, a non-saved method will be lost.
The method can be saved in the Method
register
.
27 File open error. An error occurred during opening of a file. The cause
may be a defective or damaged Flash Card (see
Chapter Flash Card).
25 File read error. An error occurred while opening a file from the Flash
Card. Cause for this is a damaged file (e.g. modified or
with wrong extension file). Old formats in principle are
always readable.
28 Please enter a valid file name. Opening or saving a file you must indicate the file
name (enter the corresponding field with a valid
name).
32 The file already exists.
Do you want to overwrite the existing file ?
If a file with the same name is already stored in
memory, it will be overwritten.
33 Files write error. Writing a file in the Flash Card an error occurred.
Cause of this may be a defective or damaged „ Flash
Card (see Chapter Flash Card).
34 Do you really want to delete this file? Security question before deletin
g
a file. A deleted file is
definitely lost.
Microwave-Program
Nr. Text Description/Remarks
1 The running time of all segments is zero.
Please enter a valid program or select a
predefined method.
You have not saved the mw program. With a new start
all not saved data will be lost. The Program run can be
saved in the Register Run.
4 The running time of all segments is zero.
Please enter a valid program or select a
predefined method.
You have not defined a program. You can create a
program or load it from the Flash Card (see Chapter
program description)
8 Your microwave program is longer than
2 hours.
For all applications except
pyroPREP
a check of the
length of the program is carried out. For length over 2