EasyManua.ls Logo

Milwaukee MW151 MAX - pH Measurement Procedure

Milwaukee MW151 MAX
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 MW150 & MW151 Bench Meters
8.3. MEASUREMENT
Remove the probe protective cap and place the tip approximately 4cm (1 ½”) into the
sample. It is recommended to wait for the sample and the pH probe to reach the same
temperature.
If necessary, press the RANGE/ until the display changes to the pH mode. Allow the
reading to stabilize (
stability tag to turn o).
The LCD will display:
y Measurement and temperature readings.
y Temperature compensation mode (MTC or ATC).
y Buers used (if option enabled in Setup).
y Electrode condition (if option enabled in Setup, MW151 only).
y The third LCD line displays: mV oset & slope values, time and date of measurement,
battery status. Use the  keys to scroll between them.
For best results is recommended to:
y Calibrate the probe before use and recalibrate periodically
y Keep the electrode hydrated
y Rinse the probe with the sample before use
y Soak in MA9015 Storage solution for at least 1 hour before measurement
MTCmode
When the probe is not connected the “NO T. PROBE” message is displayed. The MTC tag
and the default temperature (25 °C) with blinking temperature unit are displayed.
1. Press CAL/EDIT and use the  keys to set the temperature value manually.
2. Press GLP/ACCEPT to confirm or press ESC (or CAL/EDIT again) to exit without saving.

Related product manuals