pH 31
Note: The temperature value used for MTC can be set only when “NO T. PROBE
message is displayed.
8.4. WARNINGS & MESSAGES
The Calibration Check feature flags diagnostic messages during a calibration. As electrode
aging is usually a slow process, dierences between previous calibrations are likely due
to a temporary problem with the probe or buers.
Messages Displayed During Calibration
y WRONG BUFFER“ message is displayed blinking when the dierence between the
pHreading and selected buer value is significant.
Check if correct calibration buer has been used.
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
y WRONG OLD POINTS INCONSISTENT” is displayed if there is discrepancy between
new calibration value and old value recorded when calibrating with the same
probe in a buer of the same value.
Clear the previous calibration and calibrate with fresh buers. See Clear Calibration
section for details.
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME
pHmV
NO MTC ATC
USB
OPEN
AUTO
HOLD
SLOPE
CLEAR
CAL
1.68
4.01
6.86
7.01
9.18
10.01
12.45
CUST1
CUST2
DATE
ACCEPT
ºC
ºF
pH
mV
TIME

Related product manuals