EasyManuals Logo
Home>Mindray>Medical Equipment>35C50EB

Mindray 35C50EB User Manual

Mindray 35C50EB
71 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
VII
3. Do not disassemble the transducer to avoid the possibility
of electric shock.
4. Never immerse the transducer connector into liquids such
as water or disinfectant because the connector is not
waterproof. Immersion may cause electric shock or
malfunction.
5. The ultrasonic transducer is only for use with the
specified ultrasonic diagnostic system. Please refer the
ultrasonic diagnostic system operation manual to select
the proper transducer.
6. A transducer sheath must be installed over the transducer
before performing examination.
7. Do not use an aftermarket probe other than those
specified by Mindray. The probes may damage the system
causing a profound failure, e.g. a fire in the worst case.
CAUTION:
1. When using the transducer, wear sterile gloves to prevent
infection.
2. Be sure to use ultrasound gel. Please use the ultrasound
gel compliant with the relevant local regulations.
3. In normal diagnostic ultrasound mode, there is no danger
of a low-temperature burn; however, keeping the
transducer on the same region of the patient for a long
time may cause such a burn.
4. Do not use the carrying case for storing the transducer. If
the carrying case is used for storage, it may become a
source of infection.
5. The transducer and accessories supplied with it are not
delivered disinfected or sterilized. Sterilization (or high-
level disinfect) before use is required.
6. It is required to practice ALARA when operating
ultrasound system. Minimize the acoustic power without
compromising the quality of images.
7. Disposable components are packaged sterile and are
single-use only. Do not use if integrity of packaging
violated or if expiration date has passed. Please use the
disposable components compliant with the relevant local
regulations.
8. Please use the disinfection or sterilization solution that
recommended in this operator’s manual, otherwise
Mindray will not be liable for damage caused by other
solutions. If you have any questions, please contact
Mindray Customer Service Department or sales
representative.
9. Do not use pre-lubricated condoms as a sheath.
Lubricant may not be compatible with the transducer
material and damage may result.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray 35C50EB and is the answer not in the manual?

Mindray 35C50EB Specifications

General IconGeneral
BrandMindray
Model35C50EB
CategoryMedical Equipment
LanguageEnglish

Related product manuals