What to do if my Mindray Accutorr 3 monitor fails to start?
- DDiana SandersJul 31, 2025
If your Mindray Medical Equipment monitor fails to start, first ensure the AC mains is properly connected and the battery capacity is sufficient. If the issue persists, inspect the cables: between the keypad board and main board, the power board and power management board, and the main board and power management board, ensuring they are correctly connected. If cables are not the issue, the power board, power management board, or the main board may be defective and require replacement.