Mindray Accutorr 3 - Disassembling the Temperature Module (Optional)

Mindray Accutorr 3
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-9
6.5 Disassembling the Temperature Module (Optional)
6.5.1 Removing the Temperature Module PCBA and Temperature
Module Power Board PCBA
Unscrew the 2 M3 screws indicated below and remove the metal sheet.
Disconnect the Temperature board cable and the cable between Temperature
isolation power board and Temperature board. Unscrew the four M3×6 screws;
you can remove the Temperature module PCBA. Then unscrew the three M3×6
screws, you can remove the Temperature module power board PCBA.
6.5.2 Disassembling the Temperature On-Position Detection Board
PCBA
1. Unscrew the four M3×6 screws as indicated and remove the Temperature module
housing.
Cable between the isolation power board and Temperature board
3 screws
T
emperature
module
power board
PCBA
Screw for
metal sheet
T
emperature module PCBA
Metal sheet
4 screws
T
emperature
board cable

Table of Contents