4-8
4.2.5 Temperature Test
Required Tool:
Thermostatic oil tank(HART 7102 recommended) or Water Bath and Precision
Thermometer for reference.
Test procedure:
1. Set the temperature of the oil tank or water bath to 37°C and conduct the test
after the temperature stabilizes.
2. Set temperature unit to °C.
The monitor switches to from Predictive Mode to Monitor Mode when neither a
measurement is taken nor the probe is not replaced in the probe well within 60
seconds after it has been withdrawn.
3. In Monitor Mode, remove the Temperature probe from the probe sheath, insert a
probe cover, and place the probe into the oil tank or water bath.
4. Wait till the Temp value displayed on the monitor stabilizes. Verify that the
displayed value is ±0.1°C
of the oil bath thermostat setting or reference thermometer.
Contact our Technical Support Department if the Temperature test fails.
NOTE
Due to the different application environment and the test object in vivo and
vitro conditions, there is deviations in the measurement result. The
maximum deviation of 2.5℃ may exist in predictive Temperature
measurement by liquid bath techniques.