4-7
Follow this procedure to perform the NIBP cuff overpressure test:
1. Perform steps 1 to 2 in the 4.2.4.2 NIBP Accuracy Test.
2. Start the monitor. Within 10 s after you hear a beep, press and hold the
button to enter the Maintenance mode. Then press the button to display
the NIBP cuff overpressure test interface. In the PR parameter area, the code 520 is
displayed.
3. Check the manometer values and the monitor values. Both should be 0 mmHg.
Press the
button to set the patient category to adult ( ).
4. Raise the pressure in the metal vessel to 330 mmHg with the manometer bulb.
Then wait 10 seconds to allow the pressure to stabilize.
5. Press the
button to start the NIBP cuff overpressure test.
6 Press the button to set the patient category to pediatric ( ).
7. Raise the pressure in the metal vessel to 330 mmHg with the manometer bulb.
Then wait 10 seconds to allow the pressure to stabilize.
8. Press the
button to start the NIBP cuff overpressure test.
9. Press the button to set the patient category to neonate ( ).
10. Raise the pressure in the metal vessel to 165 mmHg with the manometer bulb.
Then wait 10 seconds to allow the pressure to stabilize.
11. Press the
button to start the NIBP cuff overpressure test.
If is displayed in the error code area, it indicates the NIBP cuff overpressure
test is passed. If
is displayed, the NIBP cuff overpressure test is failed. If the
test is failed, contact Mindray our your service personnel.
NOTE
You can use an NIBP simulator to replace the manometer bulb and the
reference manometer to perform the test.