Mindray A3 Service Manual

Mindray A3
436 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Service Manual
Anesthesia System
|
ANESTHESIA SYSTEM
â„¢
ANESTHESIA SYSTEM
â„¢
A5_A3_Srv Mnl Cover-046-001142-00.indd 1 10/11/11 10:59 AM

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Mindray A3 and is the answer not in the manual?

Mindray A3 Specifications

General IconGeneral
ModelA3
BrandMindray
Tidal Volume Range20-1500 ml
FiO2 Range21-100%
Gas SupplyO2, N2O, Air
Power Supply100-240 VAC, 50/60 Hz
Battery BackupYes
Ventilation ModesVCV, PCV, SIMV
Pressure Range5-60 cmH2O
PEEP0-30 cmH2O
AlarmsHigh/low airway pressure, high/low tidal volume, high/low minute volume, high/low respiratory rate, high/low FiO2, apnea, power failure, gas supply failure
TypeAnesthesia Workstation

Summary

Installation Guide

Assembly

Step-by-step instructions for unpacking and assembling the anesthesia machine.

Periodic Maintenance

Checklist before surgery

Routine tests required before anesthesia machine maintenance or use.

Functional Tests

Tests for breathing system leaks, display setup, and AG module accuracy.

Pneumatic Leak Tests

Procedures for testing cylinder and line pressure leaks in the pneumatic system.

Performance Verification

Tests for manual ventilation, APNEA alarm, and various ventilation modes.

Alarms and Fail safe Functions

Tests for Low FiO2, High FiO2, Peak Pressure, and Minute Volume alarms.

Electrical Tests

Procedures for testing auxiliary electrical outlets and electrical safety inspection.

Calibration

Calibration Warnings, Precautions, and Notes

Safety guidelines, cautions, and notes for calibration procedures.

System Calibration

Details on drive gas, calibration device settings, and system calibration items.

O2 Sensor Calibration

Procedures for calibrating the O2 sensor at 21% and 100% O2.

Flow Calibration (User)

User calibration procedure for flow sensors in the breathing circuit.

Flow Calibration (Service)

Service calibration for flow sensors and inspiratory valve, required after component replacement.

Pressure Calibration (Service)

Service calibration for airway pressure sensor, PEEP pressure sensor, and PEEP valve.

Pressure and Flow Zeroing (Service)

Procedures for manually zeroing pressure and flow sensors and troubleshooting failures.

Calibrate the AG Module

Procedure for calibrating the AG module with standard gas mixtures.

Repair and Troubleshooting

Technical Alarms Check

Troubleshooting startup alarms, CPU board, power board, and AG module runtime alarms.

Pneumatic Circuit System Problems

Troubleshooting guide for failures in the anesthetic gas delivery, ventilator, and scavenging systems.

Sensors and Valves Problems

Troubleshooting for sensor and valve failures using diagnostic tests.

Hardware and Electrical Problems

Troubleshooting guide for issues related to touch panel, display, power, and system switches.

Disassemble the Assemblies

Disassemble the Breathing System

Procedures for removing O2 sensor, breathing tubes, flow sensor, and absorbent canister.

Replacement Parts

Warranty

Warranty Statements

Details on warranty coverage, duration, exclusions, and conditions for A5/A3 system.