Battery Maintenance and Replacement Periodic Maintenance
3 - 8 046-001141-00 A5/A3™ Service Manual
FIGURE 3-5 Dust Filter
3.7 Battery Maintenance and Replacement
If the battery supply does not function normally, replace it as follows:
1. Open the rear cover of the anesthesia system.
2. Open the battery box of the anesthesia system.
3. Remove the old battery.
4. Install the new battery.
5. Close the battery box.
6. Close the rear cover.
7. Use only Mindray approved batteries (P/N 115-018012-00).
36