A5/A3â„¢ Service Manual 046-001141-00 2 - 19
Installation Guide Assembly
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded label. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
2.2.3.3 Drain the Vaporizer
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.
NOTE: The A5/A3 should use vaporizers with Selectatec mounting system that
are compliant to ISO 8835-4. Refer to the vaporizer manufacturer's
Instructions For Use for filling or draining the vaporizer.
2.2.4 Monitoring Products Mounting and Electrical Connection
(if available)
1. Mount the monitor (if available) according to the manufacturer's monitor assembly instructions.
NOTE: Use of other monitors and mounting hardware is the responsibility of
the installer.
2. After mounting a monitor to the A5/A3, connect it to one of the AC outlets located on the rear
of the A5/A3.
3. Turn on each monitor one at a time. Verify that the circuit breaker holds without tripping.
4. Dress each line cord neatly along the side of the anesthesia machine or tucked inside the
monitor arm. An optional cable routing kit is available. The cable routing kit contains three (3)
clips, screws, and two (2) ethernet cables. The clips attach to three (3) sets of holes on the rear
door of the A5/A3. Ethernet and power cables can be routed through the clips.