Pneumatic Circuit System Problems Repair and Troubleshooting
5 - 26 046-001141-00 A5/A3™ Service Manual
FIGURE 5-20
Before using the negative pressure ball, make sure that it is not leaky. Check if the front sealing cover 
is tightened. Then flatten the negative pressure ball to remove the gas inside. Install the gas outlet 
plug properly. Block the front gas inlet with your finger then release the ball. Ball should not visibly 
inflate for at least 30 seconds. If it does, replace the ball.
5.3.2 Gas Supplies and Drive Gas
The following table lists gas supplies and drive gas related failures.
Failure description Possible cause Recommended action
Leak
The gas supply tube is 
damaged or the seal at the 
connection is damaged. 
Replace the gas supply tube or the seal at the 
connection. 
The quick plug-in 
connector leaks.
Replace the quick plug-in connector or PU tube 
(when the PU tube is not damaged, if the tube is 
long enough, cut off a small segment of the tube 
where the quick plug-in connector is met, and 
then insert the tube into position). 
The pipeline gas supply 
inlet assembly leaks.
Check if the check valve of the pipeline gas 
supplies inlet assembly leaks in the reverse 
direction. Replace it if necessary. Check and 
replace the damaged seal of the pipeline gas 
supplies inlet assembly. If the problem persists, 
replace the pipeline gas supplies inlet assembly.
The drive gas pipeline 
leaks.
Check and repair the expiratory valve assembly as 
per the procedures described in ‘‘Leak Test of Low-
pressure Pneumatic Circuit System’’ on page 5-53, 
Pipeline pressure gauge 
shows inaccurate 
readings or no readings.
The pipeline pressure 
gauge is damaged.
Replace the pipeline pressure gauge.