Pneumatic Circuit System Problems Repair and Troubleshooting
5 - 64 046-001141-00 A5/A3™ Service Manual
FIGURE 5-68
7. Repeat steps 3 through 7 in “4 Leak test of all pipelines on the circuit adapter”. If the test fails, 
it indicates the CGO assembly is damaged. Check the seals in the CGO assembly and replace 
any damaged seals.
5.3.5 Tidal Volume
The following table lists tidal volume inaccuracy related failures.
Failure 
description Possible cause Recommended action
Inaccurate tidal 
volume
The flow sensor is not installed properly. Re-install the flow sensor.
The setting of fresh gas flow is 
inappropriate.
Adjust the fresh gas flow.
There are significant leaks in the 
breathing system and the fresh gas flow 
is too low.
Repair the leaking points after checking as 
per the procedures described in 
“Anesthetic Gas Delivery System” on 
page 5-35. and “Breathing System” on 
page 5-50.
* There is water build-up inside the flow 
sensor.
Remove the flow sensor and clear its inside 
water build-up.
*The membrane of the flow sensor 
assembly is distorted, dirty or its inside 
resistance changes. Zero drift occurs to 
the pressure sensor on the ventilator 
control board.
Enter the service mode and calibrate the 
flow sensor as described in “If 
measurement deviations are not corrected 
after multiple flow sensor calibrations, the 
user is recommended to replace the flow 
sensor and then perform calibration. If the 
problem persists, factory maintenance is 
necessary.” on page 4-18..
*The flow sensor pressure sampling 
pipeline is leaky.
Repair the leaking points after checking as 
per the procedures described in “Leak Test 
of Flow Sensor Pressure Sampling Pipeline” 
on page 5-51..