Cleaning and Disinfection 
5-4
 
 
Glutaraldehyde-based sterilization solution   
Trade Name  Chemical Name  Procedures 
Cidex Activated 
Glutaraldehyde 
Solution 
(applicable for CE 
region) 
2.2-2.7% 
Glutaraldehyde 
Soak the transducer into the activated 
solution for 10 hours (20-25℃). 
Please refer to the instructions provided 
by the manufacturer of the solution for 
details. 
Cidex Activated 
Dialdehyde 
Solution 
(applicable for 
FDA region) 
2.4% Glutaraldehyde 
 
Refer to the instructions provided by the chemical manufacturer concerning 
concentration of the sterilization solution, method of sterilization and dilution and 
cautions during use. 
 
Do not soak the transducer connector or the cable near it into water or any 
solution.   
 
Follow local regulations when selecting and using the sterilization solution. 
3.  Rinse the transducer with plenty of sterile water (about 2 gallons) for at least 1 minute 
to remove all chemical residues on it. Or, follow the rinsing method recommended by 
the sterilization solution manufacturer to rinse the transducer.   
4.  Wipe off the water on the transducer with sterile cloth or gauze after rinsing it. Do not 
dry the transducer by heating. 
 
Immerse the intra-operative transducer in the solution (for reference) 
 
Before safety and performance is affected, intra-operative probes can be sterilized by 
Cidex Activated Glutaraldehyde Solution for at least 217 times (10 hours for one time). 
Before safety and performance is affected, intra-operative probes can be sterilized by 
Minncare COLD STERILANT for at least 135 times (11 hours for one time). 
CAUTION:
Repeated disinfection will eventually damage the transducer, 
please check the transducer’s performance periodically.