3 System Description 3-11
4
When replacing the optical components, take care to protect the components.
Do not touch the surface of the optical elements with bare hand.
5
Wear antistatic gloves to dismount and mount the analogue board;
6
Do not drag the cable with excessive force. If necessary, you can disconnect the
connection of some cables in advance.
7
When replacing the temperature sensor, please take care to dismount the
insulation cotton, otherwise damage might occur; when reinstalling, use double
glue to stick the insulation back.
3.2.4.4 Replacement of the Temperature Measurement Assembly
The analogue board and the temperature sensor are components for measuring the
temperature on flow cell base assembly, so they should be replaced together.
Replacement of the two should be followed with parameter configuration. The
method is indicated as follows:
1 When the instrument is reinstalled, power on the system.
2 Click Maintenance on the main screen of the instrument. Select Maintenance
tab. Click Advanced and enter the password analyzer” to enter the advanced
adjustment interface.
3 Select Temp. to enter the following screen. Select Cell. Enter the A, B, C
parameters on the flow cell accessory into their respective positions, making sure
that they are in accordance with those on the accessory. Click Save.
Figure 3-11 Parameter configuration
4 Select Cuvette. Repeat the operation in step 3 to configure the A, B and C
parameters.
5 After above operations are completed, click Return. Click View para. and make
sure the inquired A, B and C parameters are in accordance with those on the
accessory. Click Return.
6 Exit the Maitenance screen and the system will restart. Follow the instruction to
complete the configuration of the temperature.

Table of Contents

Other manuals for Mindray BA-88A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BA-88A and is the answer not in the manual?

Mindray BA-88A Specifications

General IconGeneral
Light sourceHalogen lamp
Display7 inch touch screen
PrinterThermal printer
InterfaceRS232
Wavelengths340, 405, 546, 578, 630nm
Carry-over< 1%
Power SupplyAC 100~240V, 50/60Hz
Operating Temperature15-30°C

Related product manuals