10 - 6 BeneVision N Series Patient Monitor Operator’s Manual
IBP: PA IBP-S 
(mmHg)
SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25 3 to 120
IBP-D 
(mmHg
Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25 3 to 120
IBP-M 
(mmHg)
Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 120
IBP: CPP CPP 
(mmHg)
CPP × 0.68 + 8 (CPP-15) or 
35, 
(whichever is 
greater)
CPP × 0.86 + 
35
(CPP+15) or 
95, 
(whichever is 
smaller)
Adult: 20 to 235
Pediatric: 25 to175
Neonate: 25 to 100
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4 
(Venous 
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40
CO
2
EtCO
2
 
(mmHg)
0 to 32: 
remains the 
same
0 to 32: 
remains the 
same
0 to 32: 
remains the 
same
0 to 32: 
remains the 
same
Same as the 
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the 
measurement range
36 to 45: 
(EtCO
2
 - 6)
36 to 45: 
(EtCO
2
 - 6)
36 to 45: 
(EtCO
2
 + 6)
36 to 45: 
(EtCO
2
 + 6)
Same as the 
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the 
measurement range
>48: remains 
the same
>48: remains 
the same
>48: remains 
the same
>48: remains 
the same
Same as the 
measurement range
FiCO
2
None None Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
measurement range
awRR 
(rpm)
awRR × 0.5 or 
6 (whichever 
is greater)
(awRR - 10) or 
30 (whichever 
is greater)
awRR × 1.5 or 
30 (whichever 
is smaller)
(awRR+25) or 
85 rpm 
(whichever is 
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
AG EtCO
2
Same as the CO
2
 module
FiCO
2
Same as the CO
2
 module
awRR 
(rmp)
awRR × 0.5 or 
6 (whichever 
is greater)
(awRR - 10) or 
30 (whichever 
is greater)
awRR × 1.5 or 
30 (whichever 
is smaller)
awRR+25 or 
85 (whichever 
is smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
FiAA/EtAA Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
measurement range
FiO
2
/EtCO
2
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
measurement range
FiN
2
O/
EtN
2
O
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
default alarm 
limit
Same as the 
measurement range
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate