20 - 10 BeneVision N Series Patient Monitor Operator’s Manual
The numeric area does not display ScvO2 or 
SvO2 values when the Vigilance II, Vigileo, 
EV1000, or HemoSphere monitor is connected.
1. Check that the ScvO
2
 or SvO
2
 is set to display in the Screen Setup
menu
. For more information, see 
3.11.2Displaying Parameter Numerics
and Waveforms
.
2. Check that if the  ScvO
2
 /SvO
2
  parameter switch is enabled. If not, 
enable the ScvO
2
 /SvO
2
 measurement. For more information, see 
3.11.1Switching On or Off a Parameter.
3.Check that the Vigilance II, Vigileo, EV1000, or HemoSphere monitor is 
set properly as described in 
20.5Connecting the Device.
4. Check the connection of CCO/SvO
2
 cable. For more information, see 
20.5Connecting the Device.
Problem Solution