BeneVision N Series Patient Monitor Operator’s Manual C - 11
Art-S Extreme Alarm switch Off
High limit Adult: 175 mmHg 
Pediatric: 130 mmHg 
Neonate: 95 mmHg 
Low limit Adult: 75 mmHg 
Pediatric: 60 mmHg 
Neonate: 50 mmHg 
Priority High
Alarm Outputs Off
Art-D Extreme Alarm switch Off
High limit Adult: 105 mmHg 
Pediatric: 80 mmHg 
Neonate: 65 mmHg
Low limit Adult: 35mmHg 
Pediatric: 30 mmHg 
Neonate: 15 mmHg 
Priority High
Alarm Outputs Off
Art-M Extreme Alarm switch Off
High limit Adult: 125 mmHg 
Pediatric: 100 mmHg 
Neonate: 75 mmHg
Low limit Adult: 55 mmHg 
Pediatric: 40 mmHg 
Neonate: 30 mmHg 
Priority High
Alarm Outputs Off
CPP Alarm switch On
High limit Adult: 130 mmHg 
Pediatric: 100 mmHg 
Neonate: 90 mmHg 
Low limit Adult: 50 mmHg 
Pediatric: 40 mmHg 
Neonate: 30 mmHg 
Priority Med
Alarm Outputs Off
Measure (for P1, P2) All
Measure (for P3, P4) Mean only
Sensitivity Med
Speed 25 mm/sec
Item Default Setting