18
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
To adjust the alarm properties of parameters on the N Series bedside monitor:
a. Select the Alarm Setup 
 quick key.
Or
a. Select the Main Menu 
 quick key, and select Limits from the Alarm 
column.
b.  Select parameter and select
 • On/Off – to enable or disable parameter's alarm responses.
 • Select High or Low limit and adjust value using the drop down value list
 •  Priority– to adjust the audio and visual alarm indicators. Level options 
include High (Red), Medium (Yellow), Low (Yellow) and Prompt (audio 
absent).
 • Alarm Output – to activate an automatic recording/printing when the  
  alarm occurs.
c. Select the Main Menu 
 quick key or   to return to the main display.
Changing Arrhythmia Alarm Settings         
To adjust a patient’s arrhythmia alarm on the N Series bedside monitor: 
a. Select Alarm Setup 
 quick key, ECG waveform/numeric tile or Main 
Menu 
 quick key.
b. Select Arrhythmia tab.
c.  Scroll and select individual arrhythmia.
d.  Select
 • On/Off – to enable or disable arrhythmia's alarm responses. 
 •  Priority– to adjust the audio and visual alarm indicators. Level options 
include High (Red), Medium (Yellow), Low (Yellow) and Prompt (audio 
absent).
Note: If alarm is locked, alarm cannot be turned off or priority altered. 
 • Alarm Output – to activate an automatic recording when the alarm  
  occurs.
e. Select the Main Dialog 
 quick key or   to return to the main display.
To adjust a patient arrhythmia threshold setting on the N Series bedside monitor: