Probes and Biopsy 13-11
NOTE:
1.
Observe the graph here carefully to immerse the transducer. Only soak parts of
the transducer below the strain relief.
2.
Repeated disinfection will eventually damage the probe, please check the probe
performance periodically.
Compatible Disinfectants
For details about compatible disinfectants, please refer to the Cleaning & Disinfection Guide.
For details about P7-3TE disinfection, please refer to the P7-3TE accompanying manual.
13.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and stored in a
suitable place.
1. To prevent the probe from being damaged, DO NOT store it in locations where it may be
exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport the probe under the following ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure: 700 hPa to 1060 hPa
Store and transport the D7-2E probe under the following ambient conditions:
Ambient temperature: -10°C to 60°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure:700 hPa to 1060 hPa
Store and transport the DE10-3E/DE11-3E probe under the following ambient conditions:
Ambient temperature: -10°C to 50°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure:700 hPa to 1060 hPa
Store and transport the P7-3TE probe under the following ambient conditions:
Ambient temperature: -10°C to 45°C
Relative humidity: 30% to 90% (no condensation)
Atmospheric pressure:700 hPa to 1060 hPa
Store and transport the SP5-1E/C6-2GE/LM14-6E/L9-3E/V11-3HE probe under the following
ambient conditions:
Ambient temperature: -20°C to 55°C
Relative humidity: 20% to 95% (no condensation)
Atmospheric pressure:700 hPa to 1060 hPa
Store and transport the L16-4HE probe under the following ambient conditions:
Ambient temperature: 0°C to 60°C
Relative humidity: 30% to 95% (no condensation)
Atmospheric pressure:700 hPa to 1060 hPa
Store and transport the SC5-1E probe under the following ambient conditions:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80 EXP and is the answer not in the manual?

Mindray DC-80 EXP Specifications

General IconGeneral
TouchscreenYes
Transducer Ports4
Power Supply100-240V AC, 50/60 Hz
Imaging ModesColor Doppler, Power Doppler
Operating SystemWindows
Storage1 TB HDD
ConnectivityDICOM, Ethernet, USB
ApplicationsGeneral Imaging, Cardiology, Vascular, Pediatrics
Weight85 kg